Fully Automated Urine Analyzer
AUTION MAX AX-4060
AX-4060 combines a compact 530 × 530 mm footprint with comprehensive, high-accuracy urine testing, fully automated processing of up to 100 samples with STAT capability, and low-volume testing down to 1 mL plus flexible dual-feeder test strip operation.
Comprehensive, high-accuracy and multifunctional testing in a compact footprint
In addition to 11 urine chemistry parameters plus specific gravity, it also supports color measurement, turbidity measurement, and abnormal color detection—enabling comprehensive assessment in a compact 530 × 530 mm design.
Fully automated operation with high throughput
Load up to 100 samples at once and simply press START for fully automated processing, while the STAT port/STAT rack enables smooth handling of urgent samples.
Low-volume sampling and flexible test strip operation
The precise dispensing method and the micro-volume mode support low sample volumes (minimum 1 mL), and two built-in test strip feeders allow two different strip types to be loaded (or up to 400 consecutive tests when the same strip type is installed in both).
Features
Specification
| Measurement object | Urine |
| Reagent | AUTION Sticks/ Uriflet S |
| Measurement items | GLU (Glucose), PRO (Protein), BIL (Bilirubin), URO (Urobilinogen), PH (pH), BLD (Blood), KET (Ketones), NIT (Nitrite), LEU (Leukocytes), CRE (Creatinine), ALB (Microalbumin) P/C (Ratio of Protein to Creatinine), A/C (Ratio of Albumin to Creatinine) S.G. (Specific gravity), turbidity and color tone |
| Measurement range | S.G.: 1.000 to 1.050 (Resolution: 0.001) Turbidity: Three levels including “clear”, “turbid” and “dense turbid” |
| Measurement principle | Test strip: Dual-wavelength reflectance measurement (Single wavelength for BLD) S.G.: Reflection refractometry Color tone: Light-transmission measurement Turbidity: Light-scattering measurement |
| Measurement wavelength | 5 LED wavelengths (450, 525, 562, 635 and 760 nm) |
| Test strip reaction time | Approximately 60 seconds |
| Test strip storage | Two test strip storage compartments, each of which can contain different test strips |
| Test strip storage capacity | Maximum 200 test strips x two storage compartments |
| Processing speed | Normal measurement: Maximum 225 samples/hour Small volume measurement: Maximum 112 samples/hour |
| Required sample volume | Normal, STAT measurement: Minimum 2 mL Small volume measurement: Minimum 1 mL |
| Sample container | Sample tubes (length: 95 – 110 mm, diameter: 14.0 – 15.8 mm) |
| Sample loading capacity | One-way transportation (default): Maximum 50 samples Loop transportation: Maximum 100 samples |
| Warm-up time | Maximum 4 minutes |
| Display | Large color liquid crystal display (800 × 480 dots) |
| Built-in printer | 58-mm width printer paper (24 digits) |
| Memory capacity | Normal, STAT, small volume measurement: 2500 tests Control measurement: 200 tests Check measurement: 50 tests Trouble list: 100 tests |
| External output | 1 port (Replaceable with the optional Ethernet terminal) |
| Communication system | RS-232C compliant (Switchable between one-way and two-way) |
| Transmission speed | RS-232C: Selectable from 300, 600, 1200, 2400, 4800, 9600 and 19200 bps Ethernet: 10BASE-T |
| Measurement environment | Temperature: 10 – 30°C Humidity: 30 – 60% RH (No condensation) |
| Dimensions | 530(W) × 530(D) × 530(H) mm (including the sampler) |
| Weight | Main body: Approximately 40 kg Sampler: Approximately 4 kg |
| Power requirements | 100 – 240 V AC (Maximum power line fluctuation of ±10%) 50/60 Hz |
| Power input | Maximum 150 VA |